Accuracy: % Full Scale vs. % Reading

The accuracy (really inaccuracy) of mass flow instruments is specified in one of two ways, either accuracy as a percentage of full scale (% FS), or accuracy as a percentage of reading (% RD).

If an instrument has accuracy specified as % FS then the error will have a fixed value no matter where the flow is in the flow range.

Take, for example, an instrument calibrated for a flow of 100 ln/min with stated accuracy 1.0% of FS.
At a flow of 100 ln/min (full scale) the error will be 1% of full scale, or +/- 1 ln/min.
As the flow moves way from full scale the error will still be 1% FS (+/- 1 ln/min), so at a flow of 50 ln/min that error of +/- 1 ln/min becomes a larger percentage (+/- 2%) of flow.
Going further away from full scale flow further increases the error as a percentage of flow; at a flow of 10 ln/min the +/- 1 ln/min error is +/- 10% of the flow.

If, however, an instrument has accuracy specified as % RD then the error will always be the same percentage of the actual flow. 

Using the 100 ln/min instrument again as the example, but this time with a stated accuracy of 1% RD, at 10 ln/min of flow the error is only +/- 1% of the flow, better by 10 times.

What is the difference between accuracy of reading and accuracy of full scale? 

Accuracy of Reading (%RD) ensures that the error remains a consistent percentage relative to the actual flow rate. 
This means that the precision is proportional across the entire flow range.

Accuracy of Full Scale (%FS) guarantees that the error is a fixed value, unaffected by the flow’s position within the range. 
The variation can be expressed as a constant flow rate (for example, l/min), rather than as a fluctuating percentage.

Accuracy: % Full Scale vs. % Reading

MASS-VIEW Accuracy

Accuracy %RD from our MASS-VIEW calibration certificate


Preview Calibration Certificate

MASS-VIEW Calibration Certificate
MASS-VIEW Calibration Certificate
MASS-VIEW MV-106 Calibration Certificate
MASS-VIEW MV-106 Calibration Certificate

Digital Variable Area Flow instruments 

Our MASS-VIEW instruments have an Rated accuracy of ±(1.75% Rd + 0.25% FS)
Optional Calibration certificate is available for the MASS-VIEW instruments.
On the product page of the instrument you will find all the specifications: MASS-VIEW


MAG-VIEW Magnetic Flow Instruments

Most MAG-VIEW instruments have an accuracy ±0.7 % of reading (RD) and ±0.3 % of range (FS)

The low flow models 0.0083 .. 1 l/min (MVM-001-PA/PN) and 0.05 .. 2 l/min model (MVM-002-PA/PN)
Have an FS accuracy depending on the measuring range:
±1 % of 0...50 % of measuring range.
±2 % of 50…100 % of measuring range.

See the product page of the MAG-VIEW to see the accuracy of all specifications.

Mass Flow ONLINE: your online shop for flow meters!Bronkhorst-MassFlowOnline 

Mass Flow ONLINE, an online sales channel of Bronkhorst® and partners, specialized in the sales of liquid and gas flow meters. 
In the following list you will find some of the advantages that you will experience if you shop at Mass Flow ONLINE:

  • Fast delivery
  • High quality products
  • Clear competitive pricing
  • 24/7 open for business
  • Highly efficient purchase channel including order tracking
  • Easy product selection and ordering
  • Safe payment
  • Clear return policy

Discover the perfect instrument with the right accuracy here!